Cabinet fan does not turn.
Limited or Iow flow.
Low concentration.
2010-2225SP Rev -
Defective cabinet fan.
Defective electrical
connections.
Restriction in humidifier or
tubing.
Product regulator set too
low.
Leak.
Weak compressor.
Air flow obstruction.
Compressor inlet filter is
dirty or partially blocked.
System leak
Faulty compressor
Unit temperature too high,
Contaminated sieve beds.
Defective rotary valve.
Restriction in exhaust
muffler,
Restriction in suction
resonator.
SEP 04
Replace cabinet fan.
Check electrical connections.
Replace humidifier or tubing.
Adjust regulator setting.
Leak test and repair leak.
Check system pressure, and
rebuild or exchange
compressor.
Check Filter, suction
resonator, and suction tube
for obstruction.
Replace inlet filter.
Leak test and repair leak.
Check system pressure, and
rebuild or replace
compressor.
Blocked air intake or
defective cabinet filter.
Check that P1 and P2
pressures are within range.
Replace sieve module.
Repair or replace sieve
module.
Replace or clean exhaust
muffler.
Check suction resonator and
suction tube for obstruction
and remove.
Page 32 of 50