Alarm Settings (Factory Setting) - Dräger X-am 8000 Instrucciones De Uso

Ocultar thumbs Ver también para X-am 8000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 97

6.3 Alarm settings (factory setting)

Definitions:
Pre-acknowledgement: If the alarm condition is acknowledged (by pressing the OK button), the acoustic alarm and
the vibration are switched off. The alarm is only completely reset (LED and display) once the alarm condition no
longer exists.
Acknowledgement: If an A1 alarm condition that no longer exists is acknowledged (by pressing the OK button), all
alarm elements are reset.
If the A2 and A1 alarms are configured so that they can be acknowledged, the pre-acknowledgement or
acknowledgement of the A2 alarm pre-acknowledges the A1 alarm, or completely acknowledges this alarm if the
alarm condition no longer exists.
Explanation of the symbols:
: Feature activated
: Pre-acknowledgement
Alarms/events
A1 ↑ (increasing)
A2 ↑ (increasing)
A1 ↓ (decreasing)
A2 ↓ (decreasing)
1)2)
STEL
3)
TWA
4)
Error
5)
Battery pre-alarm
6)
Battery main alarm
Device error
Channel error
1) The STEL alarm can be triggered with a delay of max. 1 minute.
2) After this alarm, the deployment of personnel is subject to the relevant national regulations.
3) A TWA alarm can only be reset by turning the gas detector off and on.
4) Troubleshooting, see technical manual.
5) The battery still lasts approx. 20 minutes after the battery pre-alarm triggers.
6) The gas detector switches off automatically 20 s after a battery main alarm.
X-am
Presentation
in the display
Latching
A1
-
A2
A1
-
A2
STEL
TWA
-
-
-
-
-
-
®
8000
Acknowl-
LEDs
edgeable
-
-
-
-
-
-
Device settings
en
Horn
Vibration
-
-

Publicidad

Tabla de contenido
loading

Tabla de contenido