o
I
(
PeratIng
nstrUctIons
cont
3. To restore normal operation, remove AC power to the unit for ten (10) seconds and
then restore.
If the E04 error occurs be sure the load is within the maximum specification and properly
balanced (centered on tray) and/or reduce sample size/speed before restarting the
unit. If the E04 error occurs due to acceptable sample vibration or another vibration
source, the vibration sensing function can be disabled as described above.
BEEPER PREFERENCE (muting audible alarm)
To silence beeper operation (except for error codes), with the unit in standby mode,
press and hold the time on/off button and press the standby button. Release the
standby button first, and then release the on/off button. To restore normal beeper
operation, remove AC power to unit for ten (10) seconds and then restore.
SINGLE POINT CALIBRATION PROCEDURE
This procedure is used to fine tune and calibrate the Incubating Orbital Shaker at a
specific temperature setting. This process may be repeated for up to three (3) sepa-
rate set-points. If a fourth calibration set-point is entered, the first set-point entered
will be overwritten.
1. Turn unit on.
2. Set desired temperature.
3. Stabilize one (1) hour or more, measuring the temperature with a temperature
probe/thermometer.
4. Press and hold standby button, then press the temperature up arrow once. The
unit will beep two (2) times, confirming calibration mode. The display will now be
flashing.
5. Press the temperature up/down arrows until the display matches the temperature
probe/thermometer.
6. Press standby button to exit calibration mode and return to normal heating.
11
'
)
d
This process may be repeated at the same set-point, multiple times for fine tuning if
desired.
The unit will now use the biased offset for that specific temperature setting and
increase or decrease temperature accordingly to bring the temperature to set
temperature. The decimal point of the display will flash to indicate a biased off-
set is being used. All other temperature settings will use the standard internal
calibration. This offset will be stored in memory and retained until reset.
To restore unit to factory setting:
Press and hold the standby button while pressing the temperature down arrow once.
The reset will be confirmed with two (2) audible beeps. Press the standby button to exit
calibration mode and return to normal heating.
Speed Calibration Procedure:
This procedure is used to self calibrate the incubating shakers. The tray should be free
of any samples, vessels, and accessories prior to calibrating.
1. Turn unit on. Speed and time displays will be illuminated.
2. Press and hold the standby button and momentarily press the speed on/off button.
The speed display should read "CAL".
3. The unit will be run for approximately one (1) minute and automatically calibrate.
RS-232 SERIAL PORT
The RS-232 serial port provides two-way communications for data logging and unit
control.
t
s
echnIcal
ervIce
For more information or technical assistance contact your local VWR representative
or visit www.vwr.com.