VITAL SCIENTIFIC VITALAB Junior Manual Del Usuario página 136

Tabla de contenido

Publicidad

Mantenimiento
Mensaje de error
E63 arm vertical error
E72 arm init failed
E75 arm reset failed
E76 arm reset failed
E77 arm communication error
E100 discs 14V failed
E101 discs 30V failed
E102 rotor error
E103 sample disc error
E112 discs init failed
E115 rotor reset failed
E116 sample disc reset failed
E117 reag. /samp. discs
commun. error
E120 washarm/bellows pump
14V failed
E121 washarm/bellows pump
30V failed
E122 washarm error
E123 bellows pump error
6-20
Causa
Acción
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Informe al soporte técnico.
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Informe al soporte técnico.
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
R
.
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
R
.
ESET
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Informe al soporte técnico.
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Asegúrese de que el rotor de
cubetas está correctamente
colocado.
Pulse F5 (Specific Reset) o
seleccione C
+F6 S
TRL
YSTEM
.
R
ESET
Tenga en cuenta que este error
frecuentemente se produce con
otros errores. Intente solucionar
primero los restantes errores.
Entonces probablemente
desaparecerá también este error.
VITAL SCIENTIFIC N.V.

Publicidad

Tabla de contenido
loading

Tabla de contenido