of some people can have an oversensitive or allergic
reaction to the heat treatment (e.g. intense redness,
blisters, itching, heavy sweating).
Starting the unit
Important
• Always close or cover your eyes when radiating the
face.
• Press the On/Off switch (3)
• The unit will turn on automatically after a short time.
• Set the time selector (5) to the desired duration (1 to
15 minutes).
• You can press the time selector (5) during operation
to reset the treatment duration.
• The infrared radiant heater switches off automatically
after the set time.
• For an additional treatment, you only need to press
the time selector (5) to reset your individual treatment
time.
Important
• Do not touch the hot glass.
• To end treatment, use the On/Off switch (3) and dis-
connect the power plug.
During treatment
• Examine the part of the body treated frequently and
check for skin reactions. Cease exposure immedi-
ately if there are any signs of oversensitivity or an
allergic reaction, and consult a doctor.
• The unit can be switched off at any time using the
On/Off switch (3).
• Do not cover up or pack the unit away while it is
warm.
After treatment
• Switch off the unit at the On/Off switch (3).
• Pull the mains plug out of the socket.
6. Cleaning, changing the infrared
lamp, fuse and storage
Cleaning
Important
Before cleaning the unit, switch it off, unplug it and
allow it to cool down completely.
Important
• Ensure that no water enters the unit!
• Do not wash the unit in a dishwasher!
The unit can be cleaned with a damp cloth. Do not use
cleaning agents.
Replacing the fuse (UK plug only)
If the fuse needs replacing, please observe the
following:
• Remove the mains plug
• Lift out the fuse holder from the inside of the mains
plug using a flat screwdriver
• Pull out the fuse from the fuse holder and replace it
with a new fuse type: 3A (breaking capacity 6 kA at
240 V) in accordance with BS-1362 / SS167.
• Press the fuse holder containing the new fuse into
the plug slot
• Plug the mains plug into the socket,press ON/OFF
switch (3) and test to see if it works.
Changing the infrared lamp
Remember that a faulty lamp should only be replaced
with a lamp of the same type.
Warning
Danger of an electric shock!
• Before changing the lamp, make sure the power ca-
ble is unplugged from the socket.
• Risk of burning on hot surfaces!
• Allow the unit to cool completely before any repairs.
• Lamps are not covered by the guarantee.
Procedure
1. Place the cool unit with the glass ceramic plate down
on a flat surface.
2. Using a suitable crosshead screwdriver, undo the
8 screws in the back panel.
11