9 Control

9.1 General

The control program gives the possibility of storage and statistical evaluation of
controls measured on the instrument.
For all 130 tests programmed in the unit, 2 controls can be defined. The result of
controls for 130 tests of 365 days could be stored automatically.

9.2 Program controls

The control parameters could be programmed in the menu "parameter modify", see
the screen 2.15. The setting was described in paragraph 6.4.4

9.3 Measurement of controls

A control can be run at any time during a test run, after the test has been cali-
brated.
Select "Controls" in the "test procedure" menu, a window will be open as following:
(Screen 9.1)
1. Choice "1" if only measuring 1 control, and choice "2" if 2 controls will be meas-
ured. Select "Enter" to confirm the choice.
2. Instruction window will display: Control "#"
Hold the control requested by analyser under the sample tube and press Sample
button. The programmed volume will be aspirated and measured. The process and
result of the measurement will be displayed.
3. The result will be added to the statistical memories automatically. The evaluation
of control results can be accessed through the main menu and described in 4.4.

9.4 List controls

Select "Control" in main menu, to list the statistics of controls. The list screen (Fig
9.2) will be displayed:
1. Select the test, and the month will be checked. Click "Enter".
2. The statistic of the controls will be displayed on the screen 9.3
3. Print the result by selecting the "Print".
4. Select "Exit" to go back to the above menu.

10 General Specifications

Weight:
Dimension:
Power supply:
Power consumption:
Fuses:
Operating condition:
Storage condition:
Display:
Measurement range:
Resolving power:
Interference filters: (nm)
Half band width:
Lamp:
Temperature control: (Peltier)
Flow cell:
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
10 kg
450mm(L)x330mm(W)x140mm(H)
90-250VAC, 50/60Hz
150 W max
3.15A, slow-blow type 5x20mm
15ºC to 32ºC; RH 85% max
-20ºC to 50ºC; RH 85% max
7" color LCD (640x240, 256 colors)
0.000-2.500 abs
0.001 abs (display),
0.0001 abs (calculation)
340, 405, 510, 546, 578, 620, 670, 492
<10nm
Tungsten halogen lamp 6V/10W
25, 30, 37 ±0.1ºC and ambient temperature.
30ul quartz flow cell.
Tel 34 937 700 877 Fax 34 937 702 362