APPENDIX II EMC
The equipment meets the requirements of IEC 60601-1-2:2014.
Table 1
Guidance and manufacturer's declaration-electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of the Fingertip Oximeter
should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emission
IEC 61000-3-3
Compliance
Group 1
Class B
N/A
N/A
Electromagnetic
environment-guidance
The Fingertip Oximeter uses RF
energy only for its internal func-
tion. Therefore, its RF emissions
are very low and are not likely to
cause any interference in nearby
electronic equipment.
The Fingertip Oximeter suitable
for use in all establishments,
including domestic establishments
and those directly network that
supplies buildings used for
domestic purposes.
98