STORAGE

Keep the device and accessories dry. Avoid exposure to
direct sunlight, high moisture, and dust.
s NOTICE: If the device is to be stored long-term,
remove the battery.
s NOTICE: The battery's storage environment must
be between -4°F and 95°F (-20°C and 35°C).
s NOTICE: The device's storage environment must
be between -13°F and 140°F (-25°C and 60°C)
and 10% - 95% RH.
6700-INS-LAB-RevE18

SAFETY PROTECTIONS

The nebulizer is equipped with protection against several
unsafe situations. If these situations occur, the nebulizer
will either (1) not power up or (2) turn off automatically,
as described below.

Protection against incorrect Power Supply

If an AC/DC adapter is used to power the device, ensure
the adapter meets local power regulations; if not, the
device may not power up (the Power Indicator Light will
not illuminate).

Protection Against Overheating

During normal operation, the Power Indicator Light
will illuminate GREEN. However, in the two following
cases, the device will indicate irregularity by turning off
automatically. The device then requires approximately 45
minutes at rest before restarting.
Protection Against Overuse: The electronic
components of the device are designed to function
continuously for approximately 40 minutes. Operating
the device for too long will trigger the heat-protection
mechanism and the device will turn off automatically.
Lack of Liquid in the Reservoir: In the event of lack or
insufficient liquid in the reservoir, the device will turn off
automatically.
15