S A F E T Y F E AT U R E S

■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
4.1
5 1 5 3 7 0 - 0 8 1 0
-
C O V E R S H U T
O F F S W I T C H
Vialmix
is equipped with an automatic shut-off switch that prevents the
®
user from coming into contact with the rapidly moving shaking arm.The
shut-off switch is triggered in the following situations:
When the cover is open, the shut-off switch will prevent the activation
cycle from beginning. Should the START button be pressed while the
cover is open, the motor will not start; consequently, the activation
cycle will not begin.
If the cover is lifted during operation, the shut-off switch will cause the
activation cycle to stop.
The following table shows the LED message for each of the above
situations, the associated Vialmix
and the action required of the user:
Associated
LED
Message
Emits a brief
"CO" for 5
seconds
audible alarm
"EE" alterna-
Stops shaking,
ting with "CO"
emits an
audible alarm
response, likely cause of the message,
®
Likely Cause
Vialmix
®
of Message
Response
Cycle initiation
attempted while
cover is open
Opening the
cover during
the activation
cycle
Action
Required by
User
Close the cover
and re-start the
cycle.
Power off the unit.
Do not use the
D e f i n i t y
vial
®
since it was
not acceptably
a c t i v a t e d . R e t a i n
the vial and call
Technical Serv i c e s.
(See pg. 1 0 . 1 )