ND 4969-1 61-772/774 Ins.
Specifications:
Displays:
3-3/4 digit LCD with 4000 counts for both displays
Backlight:
Green illumination with auto-off after 4.5 minutes
Analog Bargraph:
41 segment LCD with measurements of 20 times per second
Polarity:
Automatic, positive implied, negative (-) polarity indication.
Overrange:
"OL" indication is displayed.
Zero:
Automatic
Measure Rate:
Samples 2 times per second, nominal.
Auto Power Off:
Approximately after 30 minutes of non-use.
Battery Life:
200 hours continuous with Alkaline (61-772)
150 hours continuous with Alkaline (61-774)
Low Battery Indication: The "
operating level.
Power Supply:
(1) 9V battery (NEDA 1604, JIS 006P, IEC 6F22)
Includes an isolated battery compartment.
Accuracy:
Stated accuracy at 23°C ±5°C, <75% R.H.
Temperature
0.1 x (specified accuracy) per °C,
Coefficient:
(0°C to 18°C, 28°C to 50°C).
Altitude:
6561.7 ft. (2000m)
Operating Environment: 32°F to 122°F (0°C to 50°C) at < 70% R.H.
Storage Environment:
-4°F to 140°F (-20°C to 60°C) at < 80% R.H. with battery
removed from meter
Jaw Opening:
Accepts a 2.0" (51mm) conductor
Dimensions:
10.6"H x 4.1"W x 1.9"D (270mmH x 103mmW x 48.5mmD)
Weight:
1.1 lbs. (500g) including battery
Accessories included: Carrying Case, Test Leads with alligator clip, (1) 9V
battery, operating instructions.
Safety:
Complies with EN 61010-1, EN 61010-2-032, UL 61010-1,
IEC 61010-2-032, IEC 61010-031 specifications.
Rated for Cat IV-600V/Cat III-1000V
C
Double Insulation
Instrument has been evaluated and complies with insulation category II (overvoltage
category II). Pollution degree 2 in accordance with IEC-644. Indoor use.
8/30/04
2:49 PM
Page 10
" is displayed when battery voltage drops below
US
N12966
Page 10