4650/4700 ASSEMBLY INSTRUCTIONS

Before Assembly (4650/4700):
®
The ProPlus
4650/4700 Wheelchair Patient Digital Platform Scale is shipped in one carton. Carefully inspect the
carton for shipping damage before unpacking. If damage is found, contact Customer Service. The following information
details what you will find inside the main carton.
Remove each assembly from the carton and unwrap packing material carefully, to prevent scratching the unit's parts.
Set the carton aside, for storage. To avoid damaging the scale parts when unpacking it, do not use a box cutter,
knife, scissors or any sharp object to open the protective inner packaging.
Parts List:
(1) Base / Platform Assembly (with 2 ramps, handle, wheels attached)
(1) User information
(1) Box Containing:
(1) Display Assembly with attached Load Cell Connector Cable
(1) Small Box Containing:
(1) Power Adapter
(4) Feet
Assembling a 4650/4700 Scale:
This scale takes only minor assembly.
1. Place scale base on a flat, level, dry floor—place upside down carefully.
2. Thread the 4 feet into the underside of the base. Each foot should be completely hand tightened and then
backed out 1-½ full turns. This allows for proper load cell operation for accuracy of the scale.
3. Carefully tilt the scale back to an upright position on the floor with care to watch the ramp—it may flip down
when the scale is tipped back. DO NOT DROP THE SCALE. Adjust the feet, to level the scale. Be certain
that the feet are NOT completely tightened into the base (see Step 2, above).
4. Open the scale to operational position.
5. Connect the load cell cable connector to the load cell port on the scale base.
6. Connect the power adapter cord to the power port in the display, if the adapter will be used. The cord will
be outside of the display, as shown in Figure 1. It is optional to run the cord through the channel in the
back of the display, if you prefer. To accomplish this, remove the 4 screws (A) in the back with a Phillips
screwdriver, remove the cover, run the power cord (B) through the channel, and then replace the cover,
secure it with the 4 screws and tighten.
7. The scale is now assembled. Plug the scale adapter into a power source and / or insert batteries—see the
batteries section / Figure 2 for more information regarding batteries.