pipes, telephone lines, or lightning rods, etc., which may cause electric shock.
After installation, the power cord must be easy reach to facilitate unplug in an emergency.
No items should cover the vents of medical refrigerators.
4.3 Preparation Before Using
1.
Remove all packaging components, including protective foam inside the product. Pay
attention not to put plastic bag in the place where the children can reach, to
prevent suffocation accidents.
2.
Check accessories inside the refrigerator, please check the accessories and document
according to the packing list.
3.
Cleaning: should clean the unit before using.
4.4 Standing
After levelling and cleaning the product, do not turn on the power first, and leave it for more
than 24 hours before turning on the power to ensure the normal operation of the product.
5 Power on For First Time
4.
When using a medical refrigerator for the first time, follow these steps:
1. In the no‐load condition, connect the power cord to a dedicated outlet of suitable size.
2. Check if the temperature in the cabinet reaches the required temperature. Observe that
the medical refrigerator‐freezer has a normal startup and shutdown for more than 24 hours.
After confirm that the equipment performance is normal, a small amount of items can be
placed in the medical refrigerator‐freezer (please store the items in batches, each time not
exceeding 1/3 volume of the box. Ensure that the product will stop and there is a normal
start and stop for more than 12 hours. You can add another item that does not exceed 1/3
volume of the box).
3. Try not to open the door during the cooling process, otherwise it will cause the
temperature to rise.
Attention:
The medical refrigerator‐freezer should be handled by a special person. Check the operation
status every day and record it (record once every 2 to 4 hours). The temperature in the medical
refrigerator and freezer will rise if there is a fault or stoppage. If it cannot be repaired in a short
time, remove the stored items and transfer them to a location that meets their temperature
requirements to avoid damage.
Before placing items in the product, it is necessary to confirm in advance that the required
temperature of the items matches the temperature range of the product, so as not to cause the
stored items damage due to the temperature of the medical refrigerator not reaching the
required temperature of the items.
Due to the inertia of refrigeration, there is a certain difference between the actual display
temperature and the actual temperature of the medical refrigerator‐freezer. This is a normal
phenomenon. All medical refrigerating and freezing boxes are storage equipment. It is forbidden
to put too many items that are too hot at one time, otherwise it will cause the compressor to
keep for a long time.
Machine, the temperature does not drop easily burn the compressor, the items must be put in