Before Making
Connections
Analog Load Cell
Connections
Electrical Connections
Before making any connections, verify the type of load cell interface. The fifth character
of the product ID indicates if the unit is analog or DigiTOL
TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE TERMINAL
AND WAIT AT LEAST 30 SECONDS BEFORE CONNECTING OR DISCONNECTING ANY
HARNESS OR CONNECTING OR DISCONNECTING ANY ULTRARES OR DIGITOL
Determine maximum cable length. The maximum cable length for analog load cell
connections to the terminal depends on the total scale resistance (TSR) of the scale
base. To calculate TSR:
Recommended Maximum Cable Length Based on TSR and Cable Gauge
TSR (Ohms)
350
87 (4-350Ω cells)
45 (8-350 Ω cells)
Chapter 2: Electrical Connections
CAUTION
Load Cell Input Resistance (Ohms)
TSR =
Number of Load Cells
24 Gauge
20 Gauge
(feet/meters)
(feet/meters)
800/244
2000/610
200/61
600/183
100/31
300/91
Before Making Connections
®
(1=Analog, 3=DigiTOL).
®
BASES.
16 Gauge
(feet/meters)
4000/1219
1000/305
500/152
(7/03) 2-1