Cleaning Fixed Connectors - EXFO EPM-50 Guia Del Usuario

Medidor de potencia/fuente de luz
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 104
Maintenance

Cleaning Fixed Connectors

Regular cleaning of connectors will help maintain optimum performance. Do not try
to disassemble the unit. Doing so would break the connector.
To clean fixed connectors:
1. Fold a lint-free wiping cloth in four to form a square.
2. Moisten the center of the lint-free wiping cloth with only one drop of isopropyl
alcohol.
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the wiping cloth, and do not use bottles
that distribute too much alcohol at a time.
3. Gently wipe the connector threads three times with the folded and moistened
section of the wiping cloth.
Isopropyl alcohol takes approximately ten seconds to evaporate.
Since isopropyl alcohol is not absolutely pure, evaporation will leave
microscopic residue. Make sure you dry the surfaces before
evaporation occurs.
4. With a dry lint-free wiping cloth, gently wipe the same surfaces three times with
a rotating movement.
5. Throw out the wiping cloths after one use.
6. Moisten a cleaning tip (2.5 mm tip) with only one drop of isopropyl alcohol.
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the cleaning tip, and do not use bottles that
distribute too much alcohol at a time.
7. Slowly insert the cleaning tip into the connector until it reaches the ferrule
inside (a slow clockwise rotating movement may help).
8. Gently turn the cleaning tip one full turn.
7
8
I
MPORTANT
I
MPORTANT
I
MPORTANT
9
16

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Els-50

Tabla de contenido