Error messages:

Press the central button to clear error messages, after which the measurement can be repeated.
The following messages may appear:
bAtt ........................... the batteries are low. Replace the batteries.
E 01 ........................... The probe did not move at all.
If this error message is repeated, turn the tonometer so that the collar faces down for a short time.
If the error message is still repeated take out the probe and replace with a new one. If the error
message continues to appear, replace the probe base.
E02 .............................The probe did not touch the eye.
The measurement was made from too far away or the probe base is dirty.
E03 ..............................Probe speed was too low.
The measurement was made from too far away or the tonometer was tilted upwards too much.
E04 ..............................Probe speed was too high.
The tonometer was tilted downwards. Be sure the groove is in the horizontal position.
E05 ...............................The hit was too "soft".
The probe hit the eyelid.
E06 ................................The hit was too "hard".
The probe hit the opening eyelid or calcification in the cornea.
E07 ................................ " Bad bounce".
The probe did not hit the central cornea.
Service procedures
Replace the batteries when the <bAtt> message appears.
Change the probe base if the probe does not move smoothly.
There are no other service procedures that can be carried out by the user. All other servicing and
repairs must be carried out by the manufacturer or certified service facilities.
The device can be cleaned with a damp cloth containing desinfectant.