Error messages:
Press the central button to clear error messages, after which the measurement can be repeated.
The following messages may appear:
•
bAtt ........................... the batteries are low. Replace the batteries.
•
E 01 ........................... The probe did not move at all.
If this error message is repeated, turn the tonometer so that the collar faces down for a short time.
If the error message is still repeated take out the probe and replace with a new one. If the error
message continues to appear, replace the probe base.
•
E02 .............................The probe did not touch the eye.
The measurement was made from too far away or the probe base is dirty.
•
E03 ..............................Probe speed was too low.
The measurement was made from too far away or the tonometer was tilted upwards too much.
•
E04 ..............................Probe speed was too high.
The tonometer was tilted downwards. Be sure the groove is in the horizontal position.
•
E05 ...............................The hit was too "soft".
The probe hit the eyelid.
•
E06 ................................The hit was too "hard".
The probe hit the opening eyelid or calcification in the cornea.
•
E07 ................................ " Bad bounce".
The probe did not hit the central cornea.
Service procedures
•
Replace the batteries when the <bAtt> message appears.
•
Change the probe base if the probe does not move smoothly.
•
There are no other service procedures that can be carried out by the user. All other servicing and
repairs must be carried out by the manufacturer or certified service facilities.
•
The device can be cleaned with a damp cloth containing desinfectant.
10