57
• SpO
Lo-Limit: SpO
2
tory default value for adult is 90% and 95% for Neonate.
• PR Hi-Limit: High limit setting of pulse rate; range: 100~240bpm. From 100 to 150,
the step is 1bpm, and from 150 to 240, the step is 5bpm. The factory default value
for adult is 120bpm and 160bpm for neonate.
• PR Lo-Limit: Low limit setting of pulse rate; range: 30~99bpm, and the step is
1bpm. The factory default value for adult is 50bpm and 60bpm for neonate.
Note: When the SpO
PR reading is higher than or equal to the preset high limit or the PR reading is lower
than or equal to the preset low limit, then the over-limit alert event will be activated,
that's, the alert sound "bibibibi..." occurs, and the corresponding reading(s) blinks.
When measured on neonate, if the SpO
alert setting for 10 seconds, then the alert sound and blinking display will be activated.
4.3.9 Help
On main menu screen, move the cursor on "Help" and press Confirm key "
then the oximeter help information screen, which shows SpO
urement tips, as shown in figure 4.12.
Help
Infant
Note:
Make sure the light
emitting and receiving ends
are aimed at each other. V
Figure 4.12 Help information
---SpO
measurement
2
low limit setting; range: 50%~99%, the step is 1%. The fac-
2
reading is lower than or equal to the preset alert setting or the
2
1/2
Baby L
Pediatric
ENGLISH
reading is lower than or equal to the preset
2
Help
Note:
Remove temperature probe
cover, insert the probe into earhole,
then press key to start measurement.
Figure 4.12 Help information
---TEMP measurement
and temperature meas-
2
2/2
",