Installation; Suitable Location; Connection To Power Supply; Connection To Tap Water And Waste - Selecta PRO-NITRO S Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80188 REV E
1
2 3 4 5
Reagents tanks
J
.P. SELECTA s.a.u.
Jun/2018

4 Installation

Once the unit is unpacked, visually check, the packing list (See 1.1) and that there
are no parts damaged during transportation. Save the packing for some days for
later sending.

4.1 Suitable location

Place the unit over a flat level and stable surface, suitable for the unit weight.
No longer that 1.5m from a water tap, waste output and power supplied plug.
To work comfortably, let a free space of 50cm on each side and not least than
20cm on the rear part.
Prepare to refill the water reservoir from the top of the unit.

4.2 Connection to power supply

Locate a power supply socket (Rated to 16A) near to the unit.
Use the power cord supplied or another of similar specification.
For safety reason the socket must have ground terminal, since the unit must work
grounded.
Before to plug on the socket check that its voltage match the voltage marked on
the unit plate.

4.3 Connection to tap water and waste.

Connect the water entry hose to a water tap (1). Connect the water output hose
from (2) to waste.
The hoses «remaining steam» (3) and «Steam generator empty» (4) and "waste"
(5) must go to waste.
Do not forget the white rubber white nuts on the hose's connections.
Press plastic connections by hand. Do not use any tool.
The tap water must have, at least, 1 bar pressure.
Note: It is critical in order that the unit works properly, the hose «Remaining
steam» (3) MUST NOT BE IMMERSED under water, or blended or blocked by any
reason.

4.4 Water for steam generator

The unit includes a 6 litres water reservoir, at the top, to store the water for the
steam generator.
As the system used to control the level of water on the steam generator works by
conductivity, water of at least 20 microSiemens must be used. Water, also, must
have a low content of calcium to increase the life of the steam generator.
Water with this conductivity could be obtained from a demineralised unit. If the
conductivity is lower, add (for 6l of water) near 50mg of salt or sodium carbonate.

4.5 Connection to reagents tanks

At the left hand side, there is a door to access reagent tanks, NaOH, fixative solu-
tion and HCl. Connect each one to the hose indicated with his name.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
Tel 34 937 700 877 Fax 34 937 702 362
Page: 36

Publicidad

Tabla de contenido
loading

Tabla de contenido