4. Press the
b) High and Low setting
1. Display shows " S-A-H" (High limit)
2. Press the
3. Display shows "H00000"
4. Press the keys
5. Press the
6. Display shows " S-A-L" (Low limit)
7. Press the
8. Display shows "L00000"
9. Press the keys
10. Press the
c) Buzzer choose.
1. Display shows "S-buz"
2. Press the
d) Alarm mode setting.
1. Display shows " S-A-S"
2. Press the
3. Select the desired mode:
A-STAB= alarm when the weight is stable.
A-NSTA= alarm without stable condition.
To exit from setting mode, press the
ERROR MESSAGES
• Err2, means initial zero point is not stable.
• Err3, means the loading is not enough during two point calibration.
• Err4, means the adjustment over modify range during one point calibration.
Modify the range is between 0.6 times and 1.4 times of the factory calibration
weight.
key to confirm and enter to High & Low setting automatically.
key to enter into.
y
to introduce the value.
key to confirm and save.
key to enter into.
y
to introduce the value.
key to confirm and save.
key to enter into.
key to enter into.
key.
6