Temperature Sensor In Heat Transfer Substance - Heidolph MR 3003 control Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

E

1.3 Temperature sensor in heat transfer substance

120
110
100
90
80
70
60
50
40
30
20
Benefits:
Rapidly reaches nominal temperature
-
Adjustable safety margin
-
Limitation:
Substance temperature will stabilize on a random level below nominal, as a function of temperature
-
in the heat transfer substance.
2.
Use a 2-station temperature sensor (Set2 mode)
2.1 Temperature sensor "substance medium" in the reactive substance, temperature sensor
"heat transfer medium" in the heat transfer substance
120
110
100
90
80
70
60
heating bath
50
40
30
20
Benefits:
Exact nominal temperature reached at ease
-
No thermal load on thermolabile substances
-
Limitation:
Reaches nominal value slowly
-
32
Temperature sensor in heat transfer medium
heating bath
medium
Time [min]
Use a 2-station temperature sensor
medium
Time [min]
safety temperature
set temperature
limit max. temperature
set temperature