1.
Check the function of the heating coil
1. First open the bottom
of the furnace
2. Check the resistance of the
heating coil like this:
If 110V, the resistance is about
8Ω
If 220V, the resistance is about
30Ω
If the resistance is infinite, that
means the heating coil is bro-
ken, then you need to change a
heating coil.
2.
Exchange
1. Take off the
screws here.
2. Take off the protecting
sleeve.
Then take off the screws on
the furnace body here.
There are also screws here,
take off them, then you can
take out the flange.
3. After take out the flan-
ge, you can take out the
furnace chamber, here is
the furnace without furn-
ace chamber.
EN
4. Take off the broken heating coil
and Winding a new heating coil
on the furnace chamber.
Don't forget the insulating rod!
This is the finished furnace cham-
ber with heating coil.
14