m A o u t p u t
The device is equipped with an analogue
output in 4-20 mA current which can be
associated to only one measuring channel
chosen by the user (SeeFig. 8 - terminal board
for the connection). The following procedure
shows the different steps to carry out to set the
parameters.
The same procedure can be repeated if,
ON THE OTHER HAND, it is decided to
associate the mA OUTPUT to measurement 2,
Rx or pH depending on the models. The value
of the mA output can be viewed in RUN mode
by pressing
and go back to viewing the
MEASUREMENT by pressing
Vi su al iz at io n o f t h e p ro b e m Vs
It is possible to view the real signal in mV that the probes send to the
instrument. During the reading of the measurements, press the ARROW LOW
button twice in succession.
After a few seconds the display automatically returns to the run time display.
.
ENGLISH
61