M04-02110701
AD conversion
Initial value: 2, 0: 120times/s, 1: 240times/s, 2:
rate
480times/s, 3: 960times/s.
Reserved
Calibration Parameters
Unit
Initial value: 1, 0-g 1-kg 2-t 3:lb
Initial value: 3bits. 0-0bit; 1-1bit; 2-2bits; 3-
Decimal
3bits; 4-4bits.
Initial value: 2 mV/V; 1:1mV/V;2:2mV/V;
vity
3:3mV/V
Initial value: 0. 0:0.001;1:0.002;2:0.005;
Resolution
3:0.010;4:0.020;5:0.050
Initial value: 10,000. Range: Full scale ≤Deci-
Full scale
mal*100000 ≤999999.
Zero cali-
When write in 1, the present weight will be zero,
bration
only allowed to write when scale is stable.
with
When read, it will return present millivolt value
weight
at zero calibration.
Gain cali-
Input weight value (≤Full scale)
bration
When read, it will be the zero millivolt of
with
loadcell.
weight
Input zero millivolt.
Zero point
Range: when loadcell is 2Mv/v, it is 0.020-
calibration
8.000mV, bridge voltage is 5V.
without
When read, it will be the zero millivolt of
weight
loadcell.
Gain cali-
Range:
bration
volt≤10.000mV-zero millivolt. When write in,
without
indicator will store. When read, it will be zero
weight(mi-
millivolt of loadcell.
llivolt
Gain
Write in gain weight value(≤full scale). Must
weight ca-
write in gain millivolt first, which proceed gain
libration
calibration by both weight value and millivolt
without
value when write the register.
weight
Reserved
M04 User Manual
2mV/V
(loadcell)
<
milli-
- 39 -