Use the minimum cycle parameters below:
Phase
Pre-wash
Enzyme
wash
Wash
Rinse
Thermal
Rinse
Dry
After automated cleaning, proceed to Inspection and
Functional Testing below.
INSPECTION AND FUNCTIONAL
TESTING
• Visually inspect the instrument and check for
damage and wear.
• Cutting edges should be free of nicks and should
have a continuous edge.
• Jaws and teeth should align properly.
• Moveable parts should have smooth movement
without excessive play.
• Locking mechanisms should fasten securely and
close easily.
• Long, thin instruments should be free of bending
and distortion.
REASSEMBLY AFTER CLEANING
To reassemble after cleaning:
1. Insert desired Femoral Rod into Guide Frame.
Recirculation
Time
(min:sec)
2:00
4:00
3:00
0:20
1:00
5:00
Water
Temp
Cold Tap
Water
Hot Tap
Water
60°C
Hot Tap
Water
82°C
95°C
7
Detergent
Type
N/A
Enzymatic
Cleaner
Neutral pH
Detergent
N/A
N/A
N/A