OPERATIONS GUIDE
ProMix® Amalgamator
INTRODUCTION:
The microprocessor controlled ProMix® Amalgamator is one of the most advanced
amalgam and encapsulated material mixer systems presently available in the
dental marketplace.
The ProMix® Amalgamation Unit provides:
a)
A Constant Mixing Speed which is automatically maintained during the com-
plete run/mix cycle.
b)
Two Preset Speeds with an LED indication of the selected Low speed (3700
rpm) or High speed (4200 rpm).
c)
Two Memory Locations (M1 and M2) which can be programmed for time and
mixing speed for your two most commonly used encapsulated materials (to
the manufacturer's specifications).
d)
A Timer to set the run cycle to a maximum of 30 seconds in 1 second incre-
ments for either selected speed.
A Digital LCD Display for convenient viewing of timing functions.
e)
f)
A Unique Capsule Holder (arm assembly) which will accommodate a large
variety of today's commonly used capsules (including those
with nozzles).
g)
A Safety Switch which automatically turns OFF the run cycle should the
ProMix® Amalgamator cover accidentally be opened.
Quiet and smooth running operation.
h)
i)
A Thermally Protected Motor to prevent any possible damage caused by
overheating.
Manual Cycle Interrupt Capability to reduce material waste should the timer
j)
inadvertently be set longer than required.
4
INDICATIONS:
For the trituration of dental alloys, capsule delivered temporary filling materials and
cements.
CONTRAINDICATIONS:
None known
WARNINGS:
1. We recommend that the ProMix® Amalgamator not be used in cases where a patient
or user of the ProMix® Amalgamator has been fitted with an implanted heart pace-
maker and has been cautioned against the use of small electrical appliances (such as
electric shavers, hair dryers, etc.)
2. The ProMix® Amalgamator is not suitable for use in the presence of a
flammable anesthetic mixture with air or nitrous oxide.
PRECAUTIONS:
1. This product is intended to be used as specifically outlined in the Directions for Use.
Any use of this product inconsistent with the Directions for Use is at the discretion
and sole responsibility of the practitioner.
2. Operate the ProMix® Amalgamator only on a hard, flat surface away from
countertop edges. Unsafe to operate in any other manner than specified by
DENTSPLY.
3. As with all dental procedures, use standard personal protective equipment
(i.e., wear face mask, eyewear or face shield, gloves and protective gown).
4. Do not place the unit on or next to a radiator or other heat source. Excessive heat
may damage the unit's electronics.
5. Do not operate the unit in a tightly confined space or corner. Place the unit where air
is free to circulate on all sides and beneath it.
6. The translucent acrylic arm cover/door of the ProMix® Amalgamator has a
safety interlock switch. Do not attempt to override the switch and/or operate the
ProMix® with this door open.
7. Should the ProMix® Amalgamator get damaged in any way, do not attempt to oper-
ate the unit unless thoroughly inspected and serviced by qualified DENTSPLY
Service personnel.
5