8. Always use the unit in conjunction with a properly grounded power source. Use only
the supplied power cord. Connect only to a power source for which this unit is rated.
9. DO NOT IMMERSE UNIT, POWER CORD OR PLUG IN WATER OR
DISINFECTANT SOLUTIONS.
See cleaning and disinfection instructions on page (10).
10. DECLARATION OF CONFORMITY - EU
402 Models conform with the Laboratory Equipment European Directive, 73/23/
EEC including amendments.
ADVERSE REACTIONS:
Electrical appliances can interfere with some pacemakers. If the user or patient has
been fitted with an implanted heart pacemaker, see Warnings.
STEP-BY-STEP INSTRUCTIONS:

INSTALLATION:

NOTE: Upon receiving the ProMix® Amalgamator system check the packaging
and parts for any damage that may have occurred in transit. If damage
is apparent, please contact your dealer immediately.
UNIT POWER-UP (INITIALIZATION):
Plug the supplied detachable power cord into the back of the unit and then into the
main power supply.
NOTE: Always use the unit in conjunction with a properly grounded power
source.
Activate the Power Switch on the back of the unit.
Upon AC power up, the unit will always be initialized as follows:
• the LCD display will read "00",
• the motor will be OFF,
• the "LOW" motor speed will be selected and the "LOW" LED indicator will
be ON.
If these three conditions are not apparent, do not use the unit. Contact qualified
DENTSPLY Service personnel.
NOTE: AC power interrupts and degradation causing improper operating
voltages to the microprocessor will always result in a
microprocessor master reset and re-initialization of the unit.
6

OPERATION:

The operation of the ProMix® Amalgamator can be controlled by the LCD display and
the seven (7) push buttons/pads located on the front panel (see Figure 1, inside Front
Cover).
"LCD" display
Indicates the set (desired) amalgamation cycle/
run time and counts down to "00" during operation.
"Up" arrow
Increases the programmed run time by seconds (to a
maximum of 30 seconds).
"Down" arrow
Decreases the programmed run time by seconds.
"High" (rabbit)
Selects the high amalgamation speed (4200 rpm)
and the associated LED indicator.
"Low" (turtle)
Selects the low amalgamation speed (3700 rpm)
and the associated LED indicator.
"Start/Stop" (I/O)
Starts and stops the amalgamation cycle.
NOTE: The two (2) "memory" locations are set at the factory to commonly used set-
tings. To change the settings to suit individual needs see "Setting Memory
Locations" (page 8).
"M1" (Memory #1) Stores and recalls #1 preprogrammed amalgamation speed
and time, even in the "power OFF" condition.
NOTE: The Memory #1 location is factory set to 15 seconds at Low speed.
"M2" (Memory #2) Stores and recalls #2 preprogrammed amalgamation speed
and time, even in the "power OFF" condition.
NOTE: The Memory #2 location is factory set to 20 seconds at High speed.