Descargar Imprimir esta página

Cleaning/Sterilization - Cuda surgical LIT102 Instrucciones De Utilizacion

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

6.2 RECHARGEABLE BATTERY MODEL
The rechargeable model lightsource differs from the standard only by the use of
rechargeable batteries. The rechargeable batteries provided are a high capacity
"RCR123A" 3 volt Lithium-ion battery. The charger is compatible only with the 3 volt
high capacity rechargeable Lithium-ion batteries. The 3 volt high capacity batteries
provided are compatible only with chargers specifically designed for them. Do not put
other types of batteries in the charger or attempt to charge the batteries with another
type of charger.
A pair of fully-discharged batteries will take approximately three hours to reach a full
charge. The indicator light on the charging cradle will show green when the power
converter powers the empty cradle. The indicator is red/orange when charging batteries
and green when batteries are fully charged. Batteries may be "topped off" to full charge
even if they are not discharged. Batteries may be left in the charging cradle when
charged. (If the cradle is not powered, the indicator will show green if a battery with even
a very low charge is inserted; the cradle must be powered for the indicator to be true.) It
is advised to keep pairs of batteries together in use and in charging so that they both
have approximately the same state of charge. Damage can result from mixing batteries
of different charge strengths either in use or in charging. Avoid directly short circuiting a
battery. The power converter operates on 100∼240 VAC, 50∼60Hz @ 0.5A.
Fully charged batteries will operate the lightsource at full brightness for approximately
1½ hours, followed by approximately ½ hour of usable gradually diminishing light.
7.
CLEANING / STERILIZATION
7.1 CLEANING
The LED Rover is a high quality optical device. Care for it appropriately.
A soft brush with mild detergent may be used to clean the device. The lens of the LED
inside the optic port may be wiped with a swab. Avoid harsh or abrasive material that
could scratch or in any way damage the optical surface.
7.2
STERILIZATION
The LED Rover is provided non-sterile. The device should be cleaned and
disinfected or sterilized prior to use, according to your institution's
requirements.
Caution
autoclave. Use of these processes will result in damage to the instrument, and
void its warranty.
Process per your institution's validated procedures and process parameters.
LIT102 CUDA
®
(English)
Do not reprocess the LED Rover using steam sterilization or
Surgical
Rev. O
Date of Revision:07/22/14
Page 6 of 40

Publicidad

loading