Storage And Transportation Guidelines; Waste Disposal; Start-Up; Assembly - bredent ecovac 14000930 Manual De Funcionamiento Y Mantenimiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40
• the ambient temperatures are in a range of 10-40°C [50-104°F]
• a relative humidity of 80% is not exceeded
• you observe the manufacturer's information relating to rotating tools
• the ecovac is switched off after use
Device fuse – changing the microfuse
In the event that it is no longer possible to switch on the device, please check that the fuse is working
correctly. Switch the device off first and remove the power cable from the mains!
Do not loosen the screws of the mains plug!
The compartment for the fuses is located in the middle between the mains plug (5) and the cable
connection of the and can be removed if the device is switched off and unplugged (use an instrument/
tool for easier removal). Check whether the metal filament of the removed fuse is broken. Then reinsert
the fuse holder (with the new fuse) into the unit. Switch the device on and check proper function – if
keys or display are not lit and/or the vacuum pump or the motor are not running, please send the unit
to bredent or an authorized dealer.
Materials and liquids
• Water and other liquids must not be allowed to penetrate the inside of the device, as this can lead
to short circuits and corrosion.
• Only use commercially-available dental materials that are processed with the ecovac.
• Do not use any inflammable or explosive liquids or mixtures!
• Do not use any strong, corrosive or abrasive cleaning agents to clean surfaces.
• Only carry the device from A to B when wearing safety shoes!

3. Storage and transportation guidelines

The ecovac mixing device must be stored in a dry area and in its original packaging. The device is also shipped
in the original packaging since damage to the device caused by a different packaging (box) will be at the
expense of the shipper.

4. Waste disposal

The device should be disposed of in accordance with the usual legal regulations for the disposal of waste
electronic equipment in the country in question (e.g.: recycling centre, Recycling plant).

5. Start-up

Check the device for any transport damage as soon as it is removed from the packaging. Report any transport
damage to bredent or an authorised dealer immediately. The ecovac mixing device may only be used by qua-
lified personnel, such as dental technicians and dentists, or by persons being instructed and supervised by
such qualified personnel.

5.1 Assembly

Wall mounting
The drilling template also supplied makes it easier to carry out accurate assembly. The drilling template must
be horizontally aligned using a spirit level. The mixing/suction connections (6) should be positioned at least
30 cm from the tabletop (mixing cup slot). After drilling the three holes (diameter of 8 mm), press the dowels
in and screw in the top two screws so that they protrude by approx. 6 mm. The bottom bracket is fastened
with the screw in the lower „hole" (10) at the device. Then the device can be attached to the wall and the
bottom bracket is screwed to the wall. Safe hold of the device must be ensured. Also correct the top two
screws, if necessary. Please ensure that the wall is stable and allows for the device to be mounted.
Assembly as a floor-mounted appliance (optional)
The stand REF 21000450 is attached to the rear of the device using the screws provided. There are two
different assembly heights available. Place the device on a level surface and ensure it has a secure position.
bredent GmbH & Co.KG · Weissenhorner Str. 2 · 89250 Senden · Germany · Tel. +49 7309 872-22 · Fax +49 7309 872-24
www.bredent.com · e-mail info@bredent.com
15

Publicidad

Tabla de contenido
loading

Productos relacionados para bredent ecovac 14000930

Tabla de contenido