6. Cleaning and maintenance

6.1 Cleaning

Do not clean the device with compressed air!
Only clean from the outside using a lightly dampened cloth with the mains plug unplugged. Ensure that elec-
tronic parts do not come into contact with moisture. Do not use any cleaning products containing solvents!

6.2 Maintenance

Clean the holder for the mixing cup at the unit, the holder of the cup and the seal as well as the prefilter accor-
ding to the level of contamination. The prefilter is located in the mixing/suction connections (6). Switch the
device off and remove the mains plug before changing the prefilter. The prefilter is unfastened by turning it in
an anti-clockwise direction with the help of a size 15 fork wrench and is then cleaned in an ultrasonic bath.
It is then screwed tight again by turning it in a clockwise direction. We recommend changing the prefilter every
1 to 2 years. The prefilter should not just be cleaned once the vacuum takes longer to build up, but at regular
intervals, depending on how much the device is used.
CAUTION
Never operate the device without a prefilter.
The O-ring on the mixing/suction connections (6) and on the mixing spiral should be kept clean in order to
guarantee an optimum vacuum. The vacuum pump does not require maintenance.

7. Troubleshooting

Fault
Device switches off!
Device does not start/
work!
LED display does not
illuminate!
No vacuum build up
Vacuum builds up
but mixing cup does
not stay in place
Device is too loud
Mixing spiral is jerky
or mixing results are
not correct!
18
Cause
Motor is overloaded
Network data and/or
connections, parameters do
not match
Microfuse faulty
Vacuum pump defective
Seals are contaminated or
porous
Drilled hole in the lid of the
mixing spiral is blocked
Prefilter blocked
Sound absorber is broken or
missing entirely
Electronic control
Mixing spiral is bent or
contaminated
bredent GmbH & Co.KG · Weissenhorner Str. 2 · 89250 Senden · Germany · Tel. +49 7309 872-22 · Fax +49 7309 872-24
www.bredent.com · e-mail info@bredent.com
Troubleshooting
Restart!
Turn device „off" and „on" again.
- Compare data with type-plate
- Insert mains plug
- Switch mains switch (5) to „on"
- Set knob (3) to the recommended mixing
time (> 0 s)
Replace microfuse T 2.0 [A]
Contact bredent and send device in if necessary
- Keep seals on the mixing spiral and mixing/
suction connections clean
- Exchange porous seals
Clean the plastic insert
Disassemble and clean prefilter
Remove sound absorber residues/install
missing sound absorber
Contact bredent and send device in if necessary
Clean or replace!