Sweep System Technology
The LINDA models with Sweep System technology contain the best you can get from
an ultrasonic washing machine. In fact, the brand new generator incorporates a special ultrasonic
oscillator
with Sweep System technology. With this technology, the output frequency of the ultrasonic
generator is modulated around a central frequency, so the transducers working at the frequency of 39
kHz are modulated at a frequency between 38 and 40 kHz. This frequency modulation offers the
following advantages:
decreases washing times;
prevents the damage of delicate parts to be cleaned;
decreases the effects of stationary waves increasing the distribution of ultrasonic energy in
the washing liquid;
improves cleaning results and facilitates the cavitation in liquids that are difficult to cavitate
with traditional ultrasonic systems.
Useful tips and suggestions
The type of detergent to use. To wash any instrument or object, you should principally use tap water
and dissolve in it a quantity of detergent or solution depending on the type of material residue to be
removed from the object itself. Only use undiluted detergent if it is explicitly specified in the product's
directions for use.
To wash many objects. Do not overfill the ultrasonic washing machine. Always place the objects in
a logical manner at the bottom of the rectangular rack, place it in the tray and proceed with the
appropriate washing. Too many items washed simultaneously reduces the effectiveness of cleaning
with ultrasonic.
When objects are very dirty. Use long washing times with the heating of the washing liquid
compatible with the object to be cleaned.
When to replace the cleaning solution. To achieve efficient cleaning, change the cleaning solution
often, especially if you use it to clean surgical instruments.
Shape and size. There are no special recommendations for the shape of the objects to be cleaned,
however avoid placing in the tank particularly heavy objects of very large dimensions. The liquid
must always cover the entire object.
Washing time. The washing time always depends on the type of object, type and amount of material
to be removed.
Device maintenance
Always disconnect the plug before carrying out any maintenance. To always ensure the safety of the
device, all you need to do is regularly ensure that the enclosure and power cord of the device are not
damaged.
Cleaning is the only maintenance normally required.
It must be carried out with the device disconnected electrically. Only use neutral detergents to clean
the inside of the tray and a soft cloth to clean the outside such as the frame and the command panel.
Do not leave dirt deposits inside the tray.
If the liquid outlet hole is obstructed by dirt residues, clean with a non-pointed flexible stick, making
sure that the rubber hose is not punctured.
Istruzioni d'uso / Instructions for use / Mode d'emploi / Gebrauchsanleitung / Istrucciones de uso
LINDA Ultrasonic Cleaner / Vasca Ultrasuoni – Rev.00
17