- After each sterilisation cycle, check that there is no water remaining on the inside and outside of
the packaging. Make sure that the flow indicator has changed to the correct color.
- Keep the devices in the sterilisation pouches away from light, moisture and any kind of contamina-
tion. Follow the manufacturer's recommendations as seen on the packaging.
- The duration for which the device is kept after sterilisation should not exceed 1 month. Label the
devices, specifying the expiration date.
After the expiration date, repeat the cleaning and sterilisation procedure.
In order to maintain the precision of the torque values, it is recommended that the torque
wrench be returned for calibration every 2,000 uses of every 2 years.
In the event of breakage, please contact your approved distributor or the Anthogyr after sales
service department directly.
All repairs must be carried out with parts and subassemblies approved by the
manufacturer.
Repairs must only be carried out by an approved distributor or by the after sales
service department in the factory.
For service or repair, the equipment must be returned complete and sterile to the after sales
service department. The equipment must be sent together with a document stating the problem
as well as the complete name and add ress of the dentist.
For any claims to be considered under the guarantee, please enclose a copy of the
invoice or the delivery note with the equipment.
For every service or repair, send the complete torque wrench.
Replacement parts are available for 7 years beyond sales discontinuation of the model.
15
VII.
CALIBRATION
VIII.
REPAIR