7. Error messages/troubleshooting

If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(Example Error no. 1)
Error No.
ERR 1
ERR 2
ERR 3
ERR 5
HI
LO
*If this or any other problem occurs repeatedly, please consult your doctor.
Possible cause(s) / Solutions
The systolic pressure was determined but
after wards the cuff pressure fell below 20
mmHg.
This situation can occur, for example, if the
tube becomes unplugged after the systolic
blood pressure has been measured. Additional
possible cause: No pulse has been detected.
Unnatural pressure impulses influence the
measurement result. Reason: The arm was
moved during the measurement (Artifact).
Inflation of the cuff takes too long. The cuff is
not correctly seated or the hose connection is
not sealed.
The difference between systolic and diastolic
is excessive. Measure again carefully follow-
ing proper cuff procedures and ensure meas-
urement under quiet conditions. Contact your
doctor if this message occurs frequently.
The cuff pressure is too high. Relax for 5 minutes
and repeat the measurement.*
The pulse is too low (less than 40). Repeat the
measurement.*