Descargar Imprimir esta página

Heraeus PALAMIX Instrucciones De Uso página 5

Ocultar thumbs Ver también para PALAMIX:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
Processing guidelines according to EN ISO 17664
Instrument
Ref.
Manufacturer
Cleaning and Disinfection
Instrument can be cleaned
automatically
Cleaning agent
Neutralisation (optional)
Thermal disinfection
Ultrasonic cleaning (optional)
Manual cleaning procedure
Sterilisation
The instrument must be sterilised?
Minimum 6 minutes, gravity –
no pre-vacuum (SAL 10
Minimum 3 minutes, three
pre-vacuum cycles (SAL 10
Components made of stainless steel, handle as instruments, medical device class I. Bearings to be oiled.
PALAMIX
66036163
Heraeus Medical GmbH
Philipp-Reis-Str. 8/13
61273 Wehrheim, Germany
www.heraeus-medical.com
 yes
 alkaline
 phosphoric acid
 at least 1 min., 90 °C
 yes
1. The feed rod must be separated from the grip part of the cement gun.
2. Firstly clean off gross soiling using a soft non-metallic brush (i. e. tooth-
brush) under running tap water (30 ± 5°C) for at least 90 seconds.
3. Loosen organic residues with 15 minutes of ultrasonic treatment.
4. By using a medium-hard toothbrush remove apparent soil especially at
windings, sliding surfaces and other hardly accessible parts under running
tap water (30 ± 5°C) for at least 60 seconds. At the same time move all
mechanical elements to detach organic soil and check visually that the parts
of the cement gun are absolutely clean.
5. While rinsing pay particular attention to flushing the inner areas.
6. Wipe down feed rod and the grip part using a lint free cloth.
 yes
steam
)
-6
steam
)
-6
cement gun
®
5
 pH neutral
 citric acid
 134°C
 134°C

Publicidad

loading

Productos relacionados para Heraeus PALAMIX