User's manual
07/2015
Agri-TR
Calibrations
This menu allows to modify some of the calibrations
(unapproved) and to enter new calibrations.
Click on "Meas. and calibrations" and choose "Calibrations".
Only unapproved calibrations ( "approved" box unticked) can be
modified.
The Agri-TR can not record more than 8 calibrations.
To modify a calibration, double-click on the box to be modified,
enter the modification and click on "PC => Agri-TR" to update the
Agri-TR memory.
To insert a new calibration :
- Click on "Insertion", enter the name of the new calibration and click
on "Validate". The name of the new calibration must not have more
than 2 consecutive characters in common with an approved
calibration.
- Enter the parameters of the new calibration (A0, A1, A2 and A3),
the range of the calibration (Mmin, Mmax), the HLW ref. and the
HLW. Entering Mmin,Mmax, HLW ref. and
- Click on "PC => Agri-TR" to transfer the new calibration.
Note :
For any modification on approved calibration, contact your
Chopin Technologies entitled repair organization.
EN
HLW is optional.