S-SCORT is a portable, 12V DC battery operated, suction device for resuscitation emergencies in Pre-
Hospital Care.
CAUTION: S-SCORT suction units are designed for hospital crash cart suction, patient transport and
emergency medical service. They are not designed or intended for use in extended procedures that
require prolonged high vacuum/low airflow applications, as is the case in wound drainage or endo-
scopic use or in any other procedure that produces high vacuum levels within an occluded system for
an extended period of time. Turn the S-SCORT suction unit off when it is not in use.
The unit receives power from its internal 12V DC battery or from the 12V system of the vehicle.
The battery is recharged by an internal, dual mode battery charger which is activated by the AC-DC
converter or direct connection to the vehicle electrical system. The charger monitors the battery,
charges the battery only when necessary, shuts down the unit if battery is low and signals battery con-
dition. A fully charged battery will power the unit for 45 minutes (±10%).
SYMBOL
S-SCORT
®
VX2 Models 2310, 2310B, 2310V, 2310BV, 2310-230, 2310B-230,
ETL listed, conforms to UL STD 2601-1, IEC 60601-1, EN 60601-1-2
©2003 SSCOR, Inc., S-SCORT, Duet, HI-D and Big Stick are registered trademarks of SSCOR, Inc.
General Description
S-SCORT VX2
Description of Symbols
S-SCORT VX2
LOCATION
Control panel
Serialized label
Control panel
Control panel
2310V-230, 2310BV-230
US Patent Pending
5
MEANING
Pump is on
Type B equipment
Battery level of charge
Green LED = Full
Green/Yellow LED = High
Yellow led = Half
Yellow/Red LED = Low
Red LED = Bettery depleted
Put back on Charge
Push On / Push Off