I Dc And Ac Voltages [V]; I Voltages With Low Impedance [V Elec] - Chauvin Arnoux C.A 5220G Manual De Empleo

Ocultar thumbs Ver también para C.A 5220G:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

A
: 10A DC and AC current ranges (via the 10A terminal)
mA
: Current ranges up to 400mA DC and AC (via the + terminal)
4 - TERMINALS
4mm diameter safety terminals
I COM : common, terminal receiving the black lead
I + : terminal receiving the red lead
I 10A : terminal receiving the red lead for the 10A DC and
AC ranges
DC and AC VOLTAGES
I Connect the leads to the multimeter and connect it in parallel to the circuit to
be tested.
I Place the selector switch on voltmeter (V) function (high impedance).
The instrument is in AC mode. Press the DC/AC button to select DC and vice
versa.
I Automatic range selection : read the measured value.
I Memorise the value, if necessary, by pressing the HOLD button.
If necessary, press the following buttons :
I RANGE to select the ranges manually
I MIN.MAX to read the min or max values
I
to switch on the backlight of the display
VOLTAGES WITH
LOW IMPEDANCE: V elec
Identical procedure (as above) except for positioning the selector switch on
V elec.
DC VOLTAGES
V DC
400 mV
Digital
0.1 mV
resolution
Bargraph
10 mV
resolution
V DC
Impedance
V elec
Impedance
Accuracy
Permitted
overload
(1) 4000 count display limited to 600V
4 V
40 V
400 V
1 mV
10 mV
100 mV
100 mV
1 V
10 V
10 MΩ
270 kΩ
± 1 % reading ± 1 count
600 V rms and 900 V peak
19
600 V
(1)
1 V
100 V

Publicidad