Operational Summary - AtriCure CCS-200 Manual De Operación Y Mantención

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 57

Operational Summary

The operation of System I and System II can be
done individually or simultaneously. Switch controls
slowly to minimize gas surge. The sequence of
operation can be changed as required. A flow of gas
can be heard in the probe during the operation—this
is normal in a functioning probe.
Operational Set-up
1. With the two probes plugged into the front of the
ACC2 console, open both cylinder valves (19)
fully. Check line pressure gauge (5) to insure
adequate line pressure.
CAUTION: It is not recommended to use the
system when the pressure is below 650 psi.
Use of the system when in the yellow zone will
result in reduced performance.
WARNING: DO NOT OPERATE SYSTEM IN
RED ZONE.
2. Turn the Freeze/Defrost Control knobs (8) to
the "defrost" position.
3. Close the Temperature Control valves (7) by
turning fully clockwise. Do not overtighten.
4. Turn the Maximum Freeze Control knobs (13)
to the "temperature control" position.
5. Turn the On/Off control knob (14) to the "on"
position.
You are now ready to start freezing.
ACC2 Operation & Maintenance Manual
Operational Summary
For controlled Temperature Freezing
a. Turn the Maximum Freeze Control knob to
"temperature control".
b. Slowly turn the Freeze/Defrost Control knob to
the "freeze position".
c. Turn the Temperature Control knob
counterclockwise to decrease temperature
(colder), clockwise to increase temperature
(warmer)—monitor the probe temperature meter
until the desired tip temperature is obtained. It
may be necessary to wait momentarily for the
temperature to stabilize before making final
adjustments.
For Maximum Freeze
Slowly turn the Maximum Freeze Control knob to
the "maximum freeze" position.
To Defrost
The probe can be defrosted or warmed above 0º C at
any time. Slowly turn the red Freeze/Defrost
Control knob to the "defrost" position.
6

Publicidad

Tabla de contenido
loading

Tabla de contenido