WaterLink SpinTouchFF Manual De Instruccion página 12

Ocultar thumbs Ver también para SpinTouchFF:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Other
About... lists the Serial Number, Firmware Version, Bluetooth MAC address,
Bluetooth Version and Test Count. The Test Count shows the number of
Settings
complete tests that have been performed over the lifetime of the meter.
Tap
Ranges Enabled allows the option of having test results that are out of the
range of the reagent system displayed in red. The default setting is on.
Select Disk Detection to display an error message and abort the test when
improper testing conditions, such as No Disk, No Disk Cover, Used Disk or
Meter Check Disk, are present that would result in inaccurate or no test
results. The default setting is OFF. Tap
RANGES
Freshwater FF104 (Code 4351)
Test
Display
Factor
Abbreviation
Alkalinity
Ammonia
Hardness
G HARD
Nitrate
NITRATE
Nitrite
NITRITE
pH
Phosphate
Saltwater FF203 (Code 4352)
Test
Display
Factor
Abbreviation
Alkalinity
Ammonia
AMMO
Calcium
Magnesium
Nitrate
NITRATE
Nitrite
NITRITE
pH
Phosphate
PHOS
Test results that are out of range of the reagent system wil be RED.
not be accurate. Go SETTINGS>Other Settings to turn the Ranges Enabled feature OFF.
Testing samples for Nitrate immediately after treatment with a chlorine neutralizer
containing sodium thiosulfate will give low results. Retest in 2-3 days.
to return to the Settings menu.
Range
ALK
0 – 250 ppm
AMMO
0.0 – 3.0 ppm
0 – 500 ppm
0 – 300 ppm
0.0 – 2.0 ppm
pH
4.5 – 10.0
PHOS
0.0 - 2.0 ppm
Range
ALK
0 – 300 ppm
0.0 – 3.0 ppm
Ca
200 - 800 ppm
Mg
500 – 2200 ppm
0 – 60 ppm
0.0 – 2.0 ppm
pH
6.5 – 10.0
0.0 – 2.0 ppm
to return to the Settings menu.
Accuracy
+/- 15%
< 2.0 ppm: +/- 0.2 ppm,
> 2.0 ppm: +/- 0.4 ppm
+/- 15%
+/- 20%
+/- 0.2 ppm
+/- 0.2
+/- 0.2 ppm
Accuracy
+/- 15%
< 1.0 ppm: +/- 0.2 ppm,
> 1.0 ppm: +/- 0.4 ppm
+/-15 %
+/- 15 %
+/- 20%
+/- 0.2 ppm
+/- 0.2
+/- 0.2 ppm
12
Method
Detection Limit
15 ppm
0.2 ppm
20 ppm
5 ppm
0.1 ppm
NA
0.2 ppm
Method
Detection Limit
15 ppm
0.2 ppm
NA
NA
5 ppm
0.1 ppm
NA
0.2 ppm
RED
test results may

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

3587

Tabla de contenido