7. Tap
to start test.
8. Results will be displayed.
9. Compare the results on the display to the values on the chart shown on the pouch. If
the results are not within the range shown on the pouch, contact Support.
10. Replace the Meter Check Disk in the foil pouch for storage.
LED Calibration
The LED Calibration is performed to set the brightness of the individual LEDs.
1. Follow the Cleaning procedure on page 21 to clean the light chamber and optic lenses.
2. Tap
to go to settings.
3. Tap Calibration.
4. Tap LED Cals.
5. Remove the Meter Check Disk (Code 1705) from the foil pouch. DO NOT remove the
black cover from the disk.
6. Insert the Meter Check Disk. Close the lid.
Start
7. Tap
to begin the meter check procedure.
8. When the calibration is complete the message "LED Calibration Complete" will appear.
9. Tap
to return to the Testing Menu and resume testing.
Angle Calibration
Angle Calibration checks the alignment of the hub and disk. The results are analyzed and
reported as pass or fail. If the measurements pass the settings will be saved. If the analysis
fails, contact Support.
If performing the LED Calibration and the Angle Calibration, the LED Calibration should be
performed before the Angle Calibration.
1. Follow the Cleaning procedure on page 21 to clean the light chamber and optic lenses.
2. Tap
to go to settings.
3. Tap Calibration.
4. Tap Angle Cals.
5. Remove the Meter Check Disk (Code 1705) from the foil pouch. DO NOT remove the
black cover from the disk.
6. Insert the Meter Check Disk. Close the lid.
Start
7. Tap
to begin the meter check procedure.
8. If the meter is performing satisfactorily, "Angle Calibration Successful" will be
displayed and the settings will be saved. If the meter is not performing satisfactorily,
"Angle Calibration Unsuccessful. Contact Support" will be displayed.
9. Tap
to return to the Testing Menu and resume testing.