DeFelsko PosiTest AT-M Manual De Instrucciones página 7

Ocultar thumbs Ver también para PosiTest AT-M:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Glue failures typically occur when the glue is improperly mixed (pg.
2) or the coated surface has not been adequately prepared (pg. 2).
Upon completion of the pull-off test, the dolly and coated surface
should be examined. In addition to pull-off force, many National
and International standards such as ASTM D4541 and ISO 4624
require the nature of the fracture to be recorded.
Cohesive fracture:
fracture occurs within a coating
layer (same coating on dolly
face and coated surface).
Adhesive fracture:
fracture occurs at the interface
between layers (coating on
dolly face differs from surface).
Glue failure: visible separation of the glue from itself,
the coating, or dolly (no coating visible on the dolly face).
Cohesive fracture
(within a layer)
Analysis of Test Results
Dolly
SUBSTRATE
Adhesive fracture
(between layers)
Dolly Face
Glue
6
Coated Surface
Glue failure
(coating/glue)

Publicidad

Tabla de contenido
loading

Tabla de contenido