MC 300™ NEBULIZER
Instructions for Use
Before use, wash your hands then check the nebulizer. Replace immediately if any damage is noticed.
Ensure these instructions and those supplied with the medication have been read.
1
2
Top
Green nozzle
cover
2
1
Mouthpiece
Bottom (Cup)
Cup
1
Open the nebulizer by rotating the top counter clockwise. Pour prescribed medication into the cup.
2
2
Maximum fill volume is 6 mL. Close the nebulizer, making sure the green nozzle cover is inside, then rotate
1
1
the top until it stops.
2
Place the mouthpiece onto the nebulizer. If a mask is required, the nebulizer should be used with the
2
masks listed in reorder number section. Refer to the instructions packaged with the mask.
3
Firmly attach one end of the tubing to the nebulizer.
1
Take care not to tip the device or medication could
spill. Attach the other end of the tubing to an air
supply set within the following range:
When the air supply is on, ensure the device is creating a mist.
4
Place the mouthpiece in your mouth. Ensure your lips are sealed around the edge. If using a mask,
position the mask lightly against the face. Inhale the aerosol breathing normally. Exhale normally through the
mouthpiece (or mask). Treatment time may be longer with facemask use. Continue treatment until sputter.
Cleaning Instructions
For home, hospital or clinic
• The exterior surfaces of the nebulizer may be wiped with a gauze lightly moistened with either distilled water
or 70% isopropyl alcohol (solvent)
• Remove the tubing from the bottom of the device, and disassemble the nebulizer into its 4 parts
(mouthpiece, green nozzle cover, top and cup)
• Cleaning of the nebulizer is recommended after each aerosol treatment
Any of the following four methods may be used:
A Sterile Water: Rinse the disassembled nebulizer parts thoroughly in sterile water. Shake to remove
excess water.
B 70% Isopropyl Alcohol: Wipe the outside of the nebulizer with a tissue dipped in 70% isopropyl alcohol.
Submerge the four parts in 70% isopropyl alcohol for 1 minute. Shake to remove any excess isopropyl
alcohol.
C Air Dry: Disassemble the nebulizer and shake out any excess drug.
3
4
Range
Flow
Pressure
Min
4 [L/min]
13 [psi]
Max (recommended)
8 [L/min]
30 [psi]
D Hand Washing: Wash the disassembled parts in a warm water and liquid dish soap solution. Rinse in
clean warm water.
• For all methods, always allow to air dry completely (or dry with a lint free cloth) before reassembling. It is
recommended the nebulizer bottom be briefly reconnected to the air supply to clear the air pathway.
NOTE:
• Regular cleaning will help prevent build-up of residual medication, which can affect aerosol generation
• Always follow accepted hospital procedures for cleaning between uses with the same patient
Tubing
Occasionally you may notice condensation on the inside of your tubing. This is normal. To remove the
condensation, connect one end of the tubing to your air supply and allow air to run through the tube without
the nebulizer attached. This will force air through the tubing, drying the inside. To clean the exterior, wipe
with a lightly moistened gauze with solvent (such as distilled water or 70% isopropyl alcohol). Ensure tubing
is dry prior to storage or use.
Intended Use
The nebulizer is intended to be used with pediatric (ages 2 years and above) and adult patients, who are
under the care of a licensed healthcare provider or physician. The device is designed to aerosolize prescribed
medication for inhalation by a patient in the hospital, clinic or home care environment. The nebulizer is a
single patient use device.
Duration of Use
The nebulizer is a single patient, disposable device for use up to 7 days. The device is provided non-sterile
and should not be reprocessed, autoclaved or boiled.
WARNING
• This is a single patient use device.
• Close supervision is required when the product is used with children. It contains parts that may
present a choking or strangulation hazard.
CAUTIONS
• Federal law restricts this device to sale by or on the order of a physician.
• For patients under the age of 5, a mask shall be used.
• Not made with BPA (Bisphenol A) or Natural Rubber Latex.
• Failure to properly clean the device may affect the performance.
Description
Part
Number
MC 300™ Nebulizer –
67150
EZ Twist Tubing
MC 300™ Nebulizer –
67750
Universal Tubing
Disposable Aerosol Mask – Small
65750
Disposable Aerosol Mask – Medium 65950
Disposable Aerosol Mask – Large
65850
105730-001 Rev B. ™/® are trademarks and registered trademarks of Trudell Medical International. Copyright © Trudell Medical International 2018. All rights
reserved. Printed in Canada. Manufactured in Canada with USA and Canadian Parts.
Case
Size
Manufactured for:
50/case
Monaghan Medical Corporation
5 Latour Avenue, Suite 1600, Plattsburgh, NY 12901 USA
50/case
Tel. 800-833-9653 • Fax 518-561-5088
Manufactured by:
50/case
Trudell Medical International
50/case
725 Third Street, London, ON,
50/case
Canada N5V 5G4

Publicidad

loading

Resumen de contenidos para Monaghan MC 300

  • Página 1 Part Case • Cleaning of the nebulizer is recommended after each aerosol treatment Number Size Manufactured for: MC 300™ Nebulizer – 67150 50/case Any of the following four methods may be used: Monaghan Medical Corporation EZ Twist Tubing 5 Latour Avenue, Suite 1600, Plattsburgh, NY 12901 USA A Sterile Water: Rinse the disassembled nebulizer parts thoroughly in sterile water.
  • Página 2 • Retire el tubo de la parte inferior del dispositivo y desmonte las 4 piezas del nebulizador (boquilla, tapa de de pieza de caja la boquilla verde, tapa y recipiente). Nebulizador MC 300™ – 67150 50/caja Tubo EZ Twist • Se recomienda limpiar el nebulizador después de cada tratamiento con aerosol.