1 General information on assembly

4
Installation Instructions for Wireless Module HW8614 Retrofit Kit CEREC AC
1
General information on assembly
These installation instructions describe how to retrofit the Höft&Wessel
Notes
wireless module (HW8614) in order to connect the CEREC AC acquisition
unit to a CEREC MC XL or CEREC 3 milling unit.
If an HW8612 wireless module is already installed on the left side (when
viewed from the front), you must uninstall first it and then re-install it on the
right side.
Turning off the unit.
DANGER
Potentially lethal shock hazard
People can be injured or electrical components of the unit destroyed.
➢ Switch off the unit prior to beginning work.
➢ Pull out the power cable.
CAUTION
Use of accessories
Accessories
For reasons of product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user assumes the risk of using non-approved accessories.
NOTICE
Order of tightening screws
Observe the order of tightening screws here.
Always start in the front at the handle and then tighten the screws towards
the back or bottom.
CAUTION
General information on cable breakage
If cable is kinked or twisted.
Kinking or twisting can cause the wires in the cable to break. The cable
must be replaced.
➢ Ensure that electrical lines do not become kinked or twisted.