1. Cylinder valve
2. Inlet connection gasket
3. Inlet connection (nut, bull nose, ask)
4. High pressure gauge
5. Low pressure gauge
6. Pressure adjusting knob
7. Outlet valve
8. Relief valve
9. Cover
10. Body
11. Cover for gauge
12. Marking

TECHNICAL DATAS

Gas N2
Colour Black
ID N
P1 bar 300
K- Class 4
P2 bar 60
Q1 m3/h 50
P2 = Outlet pressure
Q1 = Delivery ow
- The function of the pressure reducer is to reduce and stabilize the pressure of a gas; the reducer changes the
pressure at which the gas is kept in the cylinder into the pressure needed to use the gas.
- The pressure reducer has been designed so as to be used only and exclusively with the type of gas and at the
pressure which are shown by the marking.
- To try and use the pressure reducer with types of gasses and pressures other than those indicated can bedangerous.
- Incorrect use of the pressure reducer can cause serious damage. Users must be trained by specialist engineers.
- The pressure reducer must be treated as a precision instrument. Protect it from accidental knocks, dust, oil and
other sources of dirt.
- Do not use the pressure reducer if it is not in perfect working condition.
- When you draw gas, the cylinder must be placed upright and protected from falling.