Descargar Imprimir esta página

Welch Allyn LS100 Manual Del Propietário página 16

Ocultar thumbs Ver también para LS100:

Publicidad

1. Unplug the unit prior to cleaning.
2. The entire unit can be wiped down with a cloth dampened with a mild solution of detergent and water.
Wipe the unit dry with a clean, dry cloth. Be careful not to allow the plug prongs to get wet.
3. D
not plug the light back into the electrical outlet until the light is thoroughly dry
O
NOTE: D
not sterilize the unit.
O
CAUTION: D
not immerse the unit in cleaning solutions.
O
General: Qualified personnel should make electrical inspections of the unit.
Leakage Current Test: The leakage current measured from any exposed metal to ground with input
voltage of 120V, 60Hz (or 240V, 50Hz) and the ground wires securely connected shall not exceed 100
microamps.
Ground Continuity: The resistance measured from the ground pin of the plug to the exposed metal on
the unit should be .1 ohm maximum.
Repair: To locate qualified Service personnel, contact your local authorized Welch Allyn distributor, or
contact Welch Allyn directly at (315) 685-4560.
If returning a floor stand unit for service, disassemble the flexible arm from the pole to prevent damage.
1. Turn the power switch to the OFF position ("0") and unplug the cord from the outlet.
2. Using a small flat head screwdriver, CAREFULLY pull back the tabs at the top of the pole
slightly-just enough to disengage them from the black plastic insert in the pole.
3. Pull up on the bottom end of the flexible arm and remove it from the pole. Let the light head rest
down by the floor stand base.
4. Wrap the light head and place unit in box
NOTE: If the supply cord of this equipment is damaged it must only be replaced by a repair shop
appointed by the manufacturer.
1. Ziehen Sie vor einer Reinigung stets den Netzstecker.
2. Das gesamte Gerät kann mit einem Tuch befeuchtet mit einem milden Reinigumgsmittel abgewischt
werden. Trocknen Sie das Gerät mit einem sauberen trockenen Tuch Die Stifte des Steckers dürfen bei
der Reinigung nicht naß werden.
3. Der Strahler muß vor Wiederanschluß an das Stromnetz vollständig trocken sein.
HINWEIS: Den Strahler nicht sterilisieren.
VOSICHT: Den Strahler nicht in eine Reinigungslösung eintauchen..
Prüfung aufVerluststrom: Der Verluststrom von einem beliebigen freiliegenden Metallteil zu Erde darf
bei einer Eingangsspannung von 120 V, 60 Hz (oder 240 V ,50 Hz) und fest verbundenen Erdleitern 100
Mikroampere nicht übersteigen.

Publicidad

loading

Productos relacionados para Welch Allyn LS100