Sealing of transmitter
1. Select the proper gland size to fit installed cable size.
2. Mount O-ring properly and correctly and grease with gel.
3. Fill Sylgard potting kit in bottom part of casing.
4. Renew Silicagel bag to prevent condensation within transmitter, if necessary.
5. Mount the enclosure carefully and make sure not to damage the O-ring.
The self-calibration of the capacitive operation key will take 3 to 5 minutes once the MAG 8000 transmitter enclosure is
removed and mounted again. During this self-calibration the device cannot be operated via the operation key.
CAUTION
Maintain IP rating when mounting
The mains or line powered PUR cable (no shield) has to be mounted under the cable clamps. All cable glands have to be
sufficiently tightened to ensure the IP rating.
Connecting
Module Interface (Option)
Output A
Output B
The 3.6 V DC battery connector and pulse connection terminals are placed in the right side of the PCB board - see figure.
Connection for add-on interface modules is placed on the left side.
HL = Hardware lock key connection
V = Verification button
6
MAG 8000 Irrigation water meter
A5E02972853-002, 03/2014