6 Cleaning and Care
CAUTION
Please notice, that you have purchased a high quality product with the VELOCE handpiece.
Premature wear and malfunctions from improperly care and service shorten the lifetime of the product.
CAUTION
We recommend fully synthetic Quality Service Oil for care.
6.1 Go through the following cleaning steps after each treatment
1.
Removing of the bur
2.

External cleaning

3.
Internal cleaning
4.
Disinfection
5.
Lubrication with oil
6.
Sterilization
6.2 External Cleaning
The VELOCE handpiece can be brushed off under flowing water or cleaned by an alcohol solution. In case of need to keep clean the spray chan-
nels use the included wires. Disinfect the wire after using.
CAUTION
It is absolutely forbidden to place the handpiece in an ultrasonic unit. Results are malfunctions.
CAUTION
6.2 Disinfection
The VELOCE handpiece can be disinfected externally.
Only use DGHM listed chemical disinfectant for spray and wipe disinfection.
CAUTION
Don't place the handpiece in disinfectant baths. Results are mulfunctions.
CAUTION
6.3 Lubrication with Oil
Use "Quality Premium Service Oil" for lubrication. Screw the enclosed VELOCE oil adapter on the oil can. Place the handpiece with the
motorside on the adapter and press the spray button for 1-2 seconds. For the chuck use also the enclosed VELOCE oil adapter.
CAUTION
Lubricate the handpiece and the chuck after each treatment. This process significantly increases the economic lifetime.
CAUTION
6.4 Sterilization
Sterilize the VELOCE handpiece in an autoclave up to max. 135°C (2 bar) / 275°F (29 psi).
Autoclave three times with an initial vacuum for at least 4 minutes at 134°C / 273°F +/- 1
Autoclave using the gravitation method for at least 10 minutes at 134°C / 273°F +/- 1
Autoclave using the gravitation method for at least 60 minutes at 121°C / 250°F +/- 1
Follow the manufacturer's instructions for use.
The prepared instrument must be stored protected germ-free from dust in a dry, dark and cool room.
CAUTION
Observe the expiration date of the sterilized item.
No chlorine-based products and no chemical solvents or cleaning materials must be used.
CAUTION
Do not use dryer (post-heat). Do not sterilize the packing box.