Indications for use:
To be used in nitrous oxide/oxygen sedation systems for delivering to a patient a mixture of nitrous oxide and oxygen gases
with a maximum nitrous oxide concentration of 70%.
PIP+
Nasal Masks are single use only.
™
Contraindications:
Contraindications for use of nitrous oxide/oxygen inhalation may include patients with:
•
Some chronic obstructive pulmonary diseases
•
Severe emotional disturbances or drug-related dependencies
•
First trimester of pregnancy
•
Treatment with bleomycin sulfate
•
Methylenetetrahydrofolate reductase deficiency
•
Severe asthma
Whenever possible, appropriate medical specialists should be consulted before administering analgesic/anxiolytic agents to
patients with significant underlying medical conditions (e.g., severe obstructive pulmonary disease, congestive heart failure,
sickle cell disease, acute otitis media, recent tympanic membrane graft, acute severe head injury.
patient history with regard to these health issues.
American Academy of Pediatric Dentistry. Clinical Guidelines: Guidelines on Use of Nitrous Oxide for Pediatric Dental Patients. Pediatr Dent 2009;31(6):148-151. Available at:
1
"http://www.aapd.org/media/policies.asp". Accessed February 23, 2010.
Do not use this device for the administration of general anesthesia or as part of, or in conjunction with, a general anesthesia
administration system.
Warnings:
To be used only by a licensed and trained professional in the use of nitrous oxide, using titration method. Patient should always
be closely monitored during nitrous oxide use. If patient has an adverse reaction, reduce or stop the flow of nitrous oxide as
needed.
The O
flush button can be used to rapidly purge the lines of N
2
nasal hood and treat with pure oxygen from either the O
oxygen assisted manual resuscitator, or equivalent. Call for emergency assistance if rapid response is not achieved.
Cautions:
Federal (U.S.) law restricts this device to sale by or on order of a dentist or physician.
Accutron, Inc.
1733 W Parkside Lane, Phoenix, AZ 85027 USA
Made in USA
Part No. 28903 / REV C - 03/2021
All product names are trademarks of Crosstex International, Inc., a Cantel Medical Company, its affiliates or related companies, unless otherwise noted.
©2021 Crosstex International, Inc. All rights reserved.
accutron-inc.com
(800) 531-2221
O. If patient does not show signs of quick recovery, remove
2
resuscitator fitting or an auxiliary oxygen tank using a demand valve,
2
MT Promedt Consulting GmbH
Altenhofstrasse 80
D-66386 St. Ingbert/Germany
Te. +49 (0) 6894-581020
Fax +49 (0) 6894- 581021
Operator must review
1
100%
10%
0482
No expiration date required.
Not made with natural latex rubber.
EN
70°C
-40°C