CONVERTIDOR CC/CC 500W para aplicación Industrial y Ferroviaria
500W DC/DC CONVERTERS for Industrial and Railway application
Altitud de trabajo
4.1.1
Working altitude
Temperatura ambiente
4.1.2
Ambient temperature
Choques y vibraciones
4.1.3
Shocks and vibrations
Humedad relativa
4.1.4
Relative humidity
Variaciones de la tensión de alimentación
5.1.1.1
Power supply voltage variations
Interrupciones de la tensión de alimentación
5.1.1.2
Power supply interruptions
Factor de ondulación a la entrada
5.1.1.4
Input ripple factor
Conmutación de la alimentación
5.1.3
Power supply switching
Sobretensiones de alimentación
5.2
Power supply over-voltages
CEM Compatibilidad electromagnética
EMC Electromagnetic Compatibility
5.5
EN50121-3-2:2006
EN50121-4:2006
Protección inversión de polaridad de entrada
7.2.6
Input reverse polarity protection
Recubrimiento de protección del PCB
9.7
PCB protection
Lista de ensayos
12.2
Tests list
www.premium.es
Valores aplicables para los apartados de la norma EN50155: 2007
Applicable values for the different sections of the norm EN50155: 2007
Up to 1800m
Class T1 column 2: load at 100%
Class T2 column 2: load at 100% and output ripple <150mVpp
Class T3 column 2: load at 75%
Class TX column 2: load at 75% and output ripple <150mVpp
According EN61373:2010 Category 1 class B
Up to 95%
From 0.70 to 1.25 Un continuous
From 0.60 to 1.40 Un 0.1s
From 1.25 to 1.40 Un 1s without damage
With option H: Class S2 (10ms); otherwise Class S1 (without interruptions)
Up to 15% of Vin nom
Class C1 (0.6 Un during 100ms without interruptions)
1.40 Un 1s (impedance 1 ohm)
Test
Radiated
emissions
Conducted
emissions
Test
Electrostatic
discharge
Radiated
high-frequency
Fast transients
Surge
Conducted RF
Magnetic field
Pulse magnetic
field
P= Performance criteria, L= Line, P= PE (Protective Earth)
With option H: Active; otherwise by fuse
PCB conformal coated
1 Visual Inspection
2 Performance test
3 Cooling
4 Dry heat
6 Supply
7 Surge, ESD and burst susceptibility
8 RF Interferences
9 Insulation
11 Shocks and vibrations
13 Equipment stress screening: 24h at 40° C and load 100%
14 Low temperature storage
Norm
Port
IEC55011
Case
IEC55011
Input
Norm
Port
IEC61000-4-2
Case
IEC61000-4-3
X/Y/Z Axis
Input
Output
IEC61000-4-4
Signal
P
Input L to L
IEC61000-4-5
Input L to P
Input
Output
IEC61000-4-6
Signal
P
IEC61000-4-8
X/Y/Z Axis
IEC61000-4-9
X/Y/Z Axis
overvoltages
5/5
Frequency
30MHz...230MHz
40dB(µV/m) Qpk at 10m
230MHz...1GHz
47dB(µV/m) Qpk at 10m
150kHz...500kHz
99dB(µV) Qpk
500kHz...30MHz
93dB(µV) Qpk
Severity
Conditions
±8kV
Air (isolated parts)
±8kV
Contact (conductive parts)
20V/m
0.08...1.0GHz M. 80% 1kHz A
10V/m
1.4...2.1GHz M. 80% 1kHz
5V/m
2.1...2.5GHz M. 80% 1kHz
±2kV
Tr/Th: 5/50 ns
±2kV
Tr/Th: 5/50 ns
±2kV
Tr/Th: 5/50 ns
±1kV
Tr/Th: 5/50 ns
±1kV
Tr/Th: 1.2/50µs
±2kV
Tr/Th: 1.2/50µs
10V
0.15...80MHz M. 80% 1kHz
10V
0.15...80MHz M. 80% 1kHz
10V
0.15...80MHz M. 80% 1kHz
10V
0.15...80MHz M. 80% 1kHz
300A/m
0Hz, 16.7Hz, 50/60Hz
300A/m
Tr/Th: 6.4/16µs
Routine
Routine
Type
Type
Type
Type
Type
Routine
Type
Routine
Type
06-11-2013
9502-126-11
CRS-500
Limits
P
B
B
A
A
A
A
A
A
B
B
A
A
A
A
A
B
CA-439-11