Indications For Use - Braun Easypump C-bloc RA Manual Del Usuario

Prospecto de la bomba

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
Pump position - position the pump at approximately the same level as the catheter
site:
Positioning the pump above this level increases flow rate.
Positioning the pump below this level decreases flow rate.
Temperature
The Select-A-Flow* device should be worn outside clothing and kept at room
temperature.
To ensure flow rate accuracy, do not place heat or cold therapy in close proximity to
the flow controller.
Temperature will affect solution viscosity, resulting in faster or slower flow rate.
Select-A-Flow* device have been calibrated using Normal Saline (NS) as the
diluent and room temperature (22°C, 72°F) as the operating environment. Flow rate
will increase approximately 1.4% per 1°F/0.6°C increase in temperature and will
decrease approximately 1.4% per 1°F/0.6°C decrease in temperature.
If refrigerated, allow pump to reach room temperature before using.
It may take approximately 12-15 hours for a pump to reach room temperature.
(See table below)
Fill Volume (ml)
Refrigerator to Room Temp (hr)
Storage
Storage of a filled Easypump
starting infusion may result in a slower flow rate.
External pressure
External pressure such as squeezing or laying on pump increases flow rate.

INDICATIONS FOR USE

Easypump
C-bloc* RA
®
(such as local anesthetics) to or around surgical wound sites and/or close proximity
to nerves for preoperative, perioperative and postoperative regional anesthesia and/
or pain management. Routes of administration include: intraoperative site, perineural,
percutaneous and epidural.
Easypump
C-bloc* RA
®
use when used to deliver local anesthetics to or around surgical wound sites, or close
proximity to nerves, when compared to narcotic only pain management.
CONTRAINDICATIONS
Easypump
C-bloc* RA
®
emulsions, or Total Parenteral Nutrition (TPN).
Easypump
C-bloc* RA
®
270
12
C-bloc* RA
®
Pump is intended to provide continuous delivery of medication
Pump is indicated to significantly decrease pain and narcotic
Pump is not intended for blood, blood products, lipids, fat
Pump is not intended for intravascular delivery.
400
15
Pump for more than 8 hours prior to
4