Pump position - position the pump at approximately the same level as the catheter
site:
•
Positioning the pump above this level increases flow rate.
•
Positioning the pump below this level decreases flow rate.
Temperature
•
The Select-A-Flow* device should be worn outside clothing and kept at room
temperature.
•
To ensure flow rate accuracy, do not place heat or cold therapy in close proximity to
the flow controller.
•
Temperature will affect solution viscosity, resulting in faster or slower flow rate.
•
Select-A-Flow* device have been calibrated using Normal Saline (NS) as the
diluent and room temperature (22°C, 72°F) as the operating environment. Flow rate
will increase approximately 1.4% per 1°F/0.6°C increase in temperature and will
decrease approximately 1.4% per 1°F/0.6°C decrease in temperature.
•
If refrigerated, allow pump to reach room temperature before using.
•
It may take approximately 12-15 hours for a pump to reach room temperature.
(See table below)
Fill Volume (ml)
Refrigerator to Room Temp (hr)
Storage
•
Storage of a filled Easypump
starting infusion may result in a slower flow rate.
External pressure
•
External pressure such as squeezing or laying on pump increases flow rate.
INDICATIONS FOR USE
•
Easypump
C-bloc* RA
®
(such as local anesthetics) to or around surgical wound sites and/or close proximity
to nerves for preoperative, perioperative and postoperative regional anesthesia and/
or pain management. Routes of administration include: intraoperative site, perineural,
percutaneous and epidural.
•
Easypump
C-bloc* RA
®
use when used to deliver local anesthetics to or around surgical wound sites, or close
proximity to nerves, when compared to narcotic only pain management.
CONTRAINDICATIONS
•
Easypump
C-bloc* RA
®
emulsions, or Total Parenteral Nutrition (TPN).
•
Easypump
C-bloc* RA
®
270
12
C-bloc* RA
®
Pump is intended to provide continuous delivery of medication
™
Pump is indicated to significantly decrease pain and narcotic
™
Pump is not intended for blood, blood products, lipids, fat
™
Pump is not intended for intravascular delivery.
™
400
15
Pump for more than 8 hours prior to
™
4