WARNINGS
Read the instructions carefully before use. It is re commended that you consult your
insulin pump manual for information regarding pump therapy.
When using the infusion set for the first time, your doctor or healthcare team must be
present.
• This device remains sterile until the package is opened or damaged. Do not use if the
sterile package has been opened or damaged.
• Protect the product from excessive humidity, sunlight and heat. Store at room
tem perature.
• Before inserting the infusion set, prepare the infusion site, according to CDC (Centers
for Disease Control and Prevention) guidelines or institutional protocol.
• Discard your infusion set after use according to local regulations. Do not clean or
re-sterilize.
• The infusion set is for subcutaneous use only.
• Always screw the luer-lock connector firmly into the pump adapter. If this is not done,
a loose connection between the infusion set and the pump could cause leakage.
Avoid using any tools to screw the luer-lock connector into place, otherwise damage
could result to the luer-lock con nector.
• Do not leave air bubbles in the infusion set. Fill carefully.
• Never insert the infusion set before priming, or attempt to release blockages while
it is in use, otherwise there is a danger of an uncontrolled quantity of insulin being
infused into the body.
4