TECHNICAL SPECIFICATIONS

Channel
Pulse Intensity
Pulse Width
Pulse Rate
Timer
Patient Compliance
Counter
Patient Lock System
Wave Form
LCD
Function Mode
Dual, isolated between channels
Adjustable 0 mA – 80 mA peak into 500 ohm load each channel,
constant current
30uS-300uS (adjustable)
1Hz-150Hz (adjustable)
10, 20, 30, 40, 50, 60 minutes and continuous
To show the usage time by user, counted per each 30 minutes treatment
time
To prevent changes in any set parameter by physician / clinician. To
unlock or lock the unit, press the UP and DOWN buttons simultaneously
for 2-3 seconds
Mono positive-phasic square pulse, negative-phasic square pulse,
asymmetrical bi-phasic square pulse
Displays Timer, Function Mode, Alternate Mode, Pulse Width, Pulse Rate,
Low Battery, Patient Compliance Counter, and Patient Lock System
M Mode
Modulated width: Pulse width is automatically varied in an
interval of 6 seconds. The modulation range of pulse width
is from setting value to 60% less than the control setting
value, then returns to the setting value. Pulse rate and
pulse width are adjustable.
MM Mode
Modulated width and rate: This mode consists of
alternating modulated width and modulated rate.
One parameter is always decreasing while the other is
increasing and vice-versa. The modulation range is 50%
decreasing.
12